ESIB- Receiver<br />

ESIB<br />

A frequency scan can be performed after setting the start and stop frequency and the step width. The<br />

scan subranges can be defined in a table (DEFINE SCAN softkey). The entry of standard-conformal<br />

frequency ranges and associated receiver settings are facilitated by default settings (softkeys CISPR<br />

RANGE A...D).<br />

The scan is started with the SCAN softkey or the RUN key in the SWEEP field.<br />

To reduce the amount of data in RFI voltage measurements, a list of subrange maxima (softkey PEAK<br />

LIST) can be generated from the scan results and an acceptance line (softkey MARGIN) may be<br />

defined. A final measurement is performed only for frequencies with high RFI level.<br />

The final measurement may be either automatic (softkey AUTOMATIC FINAL) or in interactive mode<br />

(softkey INTERACTIVE). Automatic control functions for line impedance stabilization networks (softkey<br />

LISN) are available. The final measurement is started with the softkey RUN FINAL MEAS.<br />

The measurement examples given in section 2 serve as introduction into the different fields of<br />

application of the <strong>EMI</strong> test receiver.<br />

When the <strong>EMI</strong> receiver mode is active, the MODE key directly opens the main receiver menu. For a<br />

change to the ANALYZER mode, the main menu must be called first.<br />

Operation on a Discrete Frequency<br />

Setting the Receive Frequency<br />

CONFIGURATION MODE - <strong>EMI</strong> <strong>RECEIVER</strong> menu<br />

<strong>RECEIVER</strong><br />

FREQUENCY<br />

The <strong>RECEIVER</strong> FREQUENCY softkey activates the entry field of the receive<br />

frequency.<br />

The receive frequency can also be activated with the FREQ key in the<br />

FREQUENCY field without quitting the currently selected menu. A precondition is<br />

that the frequency and level display is active.<br />

The adjustable frequency range depends on the selected input:<br />

Input 1: 20 Hz ≤ f in ≤ f max<br />

Input 2: 20 Hz ≤ f in ≤ 1 GHz<br />

When the tuning frequency is lower than twice the IF bandwidth, the IF bandwidth<br />

is automatically reduced so that this condition is met again.<br />

If the frequency is increased again, the original IF bandwidth is restored. The<br />

memory is cleared when the IF bandwidth is manually changed.<br />

The resolution of the receive frequency is always 0.1 Hz.<br />

IEC/IEEE-bus command :[SENSe:]FREQuency[:CW|FIXed] <br />

1088.7531.12 4.76 E-15