CALCulate Subsystem<br />

ESIB<br />

:CALCulate:LIMit:CONTrol:MODE RELative | ABSolute<br />

This command selects the relative or absolute scaling for the X-axis of the selected limit line.<br />

Example:<br />

":CALC:LIM:CONT:MODE REL"<br />

Features: *RST value: ABSolute<br />

SCPI: device-specific<br />

Modes:<br />

A, VA<br />

Upon selection of RELative, the unit is switched to DB.<br />

:CALCulate:LIMit:CONTrol:UNIT[:TIME] S | SYM<br />

This command defines the unit of the x-axis scaling of limit lines.<br />

Example:<br />

":CALC:LIM:CONT:UNIT SYM"<br />

Features: *RST value: S<br />

SCPI: device-specific<br />

Mode:<br />

VA<br />

:CALCulate:LIMit:CONTrol:SHIFt <br />

This command shifts a limit line which has been specified for relative frequencies or times (X-axis).<br />

Example:<br />

":CALC:LIM:CONT:SHIF 50kHz"<br />

Features: *RST value: --<br />

SCPI: device-specific<br />

Modes:<br />

R, A, VA<br />

This command is an event which is why it is not assigned an *RST value and has no query.<br />

:CALCulate:LIMit:CONTrol:SPACing LINear | LOGarithmic<br />

This command makes a selection between linear and logarithmic interpolation for determining the<br />

limit line from the frequency points.<br />

Example:<br />

":CALC:LIM:CONT:SPAC LIN"<br />

Features: *RST value: LIN<br />

SCPI: device-specific<br />

Modes:<br />

R, A, VA<br />

:CALCulate:LIMit:UPPer[:DATA] ,..<br />

This command defines the values for the upper limit lines.<br />

Example: ":CALC:LIM:UPP -10,0,0,-10"<br />

Features: *RST value: - (LIMit:STATe is set to OFF)<br />

SCPI: conforming<br />

Modes:<br />

R, A, VA<br />

The number of values for the CONTrol axis and the corresponding UPPer limit line have to be<br />

identical. The unit must be identical with the unit selected by command CALC:LIM:UNIT. If the<br />

measured values exceed the UPPer limit line in analyzer or vector analyzer mode, the limit check<br />

signals errors. The unit must be identical with the unit selected by CALC:LIM:UNIT.<br />

The units DEG, RAD, S, and HZ are available in the vector signal analysis mode only.<br />

1088.7531.12 6.24 E-15