Display and Limit Lines - Receiver<br />

ESIB<br />

Setup of Display and Limit Lines – LINES Key Field<br />

Display Lines – D LINES Key<br />

Display lines are aids which, similar to markers, make the evaluation of measurement curve data more<br />

convenient. The function of display lines is similar to that of a movable scale which can be used to<br />

measure absolute and differential values on measurement curves.<br />

In addition, the display lines can also be used to limit the range of search for marker functions.<br />

The ESIB provides four different types of display lines:<br />

• two horizontal threshold lines for setting levels or for defining level search ranges – Display Line 1/2,<br />

• two vertical frequency lines for indicating frequencies or for determining frequency search ranges –<br />

Frequency Line 1/2,<br />

• a threshold line which, for example, sets the search threshold for maximum levels (Peak Search) –<br />

Threshold Line<br />

• a reference line which serves as the basis for mathematical correlation between measurement curves<br />

– Reference Line<br />

For purposes of clarity, each line is annotated on the right side of the display diagram by the following<br />

abbreviations:<br />

D1 Display Line 1 TH Threshold Line<br />

D2 Display Line 2 REF Reference Line<br />

F1 Frequency Line 1<br />

F2 Frequency Line 2<br />

The lines for level, threshold and reference are displayed as continuous, horizontal lines over the full<br />

width of the diagram. They are movable in the y-direction.<br />

The lines for frequency are displayed as vertical, continuous lines over the total height of the diagram.<br />

They are movable in the x-direction.<br />

Note:<br />

The softkeys for setting and switching the display lines on and off operate similar to a threeposition<br />

switch:<br />

Initial condition: The line is switched off (softkey has grey background)<br />

1 st press: The line is switched on (softkey has red background) and data entry is<br />

activated. The position of the display line can be adjusted by using the rollkey,<br />

the step keys or through direct numerical inputs via the data-entry<br />

keypad. If some other arbitrary function is requested, the data-entry keypad is<br />

disabled. In this case, the display line remains switched on (softkey has green<br />

background).<br />

2 nd press: The display line is switched off (softkey has grey background).<br />

Initial condition: The line is switched on (softkey has green background)<br />

1 st press: The line is switched on (softkey has red background) and data entry is<br />

activated. The position of the display line can be adjusted by using the rollkey,<br />

the step keys or through direct numerical inputs via the data-entry<br />

keypad. If any other arbitrary function is requested, the data-entry keypad is<br />

disabled. In this case, the display line remains switched on (softkey has green<br />

background).<br />

2 nd press: The display line is switched off (softkey has grey background).<br />