ESIB<br />

Receiver - Operation on a Discrete Frequency<br />

Setting the RF Attenuation<br />

ESIB is equipped with two signal inputs selectable in the INPUT menu. With INPUT 1 selected the RF<br />

attenuation can be set between 0 and 70 dB in 10 dB steps. With INPUT 2 - this is the pulse-resistant<br />

input - the attenuation can be set between 0 and 70 dB in 5 dB steps.<br />

CONFIGURATION MODE - <strong>EMI</strong> <strong>RECEIVER</strong> menu<br />

ATTEN<br />

ATTEN<br />

ATTEN<br />

MANUAL<br />

0 DB MIN<br />

ON OFF<br />

AUTO RANGE<br />

ON OFF<br />

AUTOPREAMP<br />

ON OFF<br />

The ATTEN softkey opens a submenu where the input<br />

attenuator can be configured. The submenu includes setting<br />

the RF attenuation and the autorange function.<br />

ATTEN<br />

MANUAL<br />

The RF ATTEN MANUAL softkey activates the attenuation<br />

entry field.<br />

The following attenuation can be set depending on the active<br />

input:<br />

• INPUT 1: 0 to 70 dB in 10 dB steps,<br />

• INPUT 2: 0 to 70 dB in 5 dB steps..<br />

Other entries are rounded to the next higher integer.<br />

IEC/IEEE-bus command<br />

:INPut:ATTenuation <br />