SENSe Subsystem<br />

ESIB<br />

:[SENSe:]SCAN:BANDwidth:RESolution f min to f max<br />

This command defines the bandwidth for a receiver subscan.<br />

Example: ":SCAN1:BAND:RES 1MHz"<br />

Features: *RST value: 9 kHz (range 1)<br />

120 kHz (range 2)<br />

SCPI: device-specific<br />

Mode:<br />

R<br />

:[SENSe:]SCAN:TIME 100 µs to 100 s<br />

This command defines the measurement time of the receiver subscan.<br />

Example: ":SCAN1:TIME 1 ms"<br />

Features: *RST value: 1 ms (range 1)<br />

100µs (range 2)<br />

SCPI: device-specific<br />

Mode:<br />

R<br />

:[SENSe:]SCAN:INPut:TYPE INPUT1 | INPUT2<br />

This command defines the input for a receiver subscan.<br />

Example: ":SCAN1:INP:TYPE INPUT2"<br />

Features: *RST value: INPUT1<br />

SCPI: device-specific<br />

Mode:<br />

R<br />

:[SENSe:]SCAN:INPut:ATTenuation dB min to dB max<br />

This command defines the RF attenuation for a receiver subscan.<br />

Example: ":SCAN1:INP:ATT 30dB"<br />

Features: *RST value: 10 dB<br />

SCPI: device-specific<br />

Mode:<br />

R<br />

:[SENSe:]SCAN:INPut:ATTenuation:AUTO ON | OFF<br />

This command switches on or off the autoranging function in a receiver subscan.<br />

Example: ":SCAN1:INP:ATT:AUTO ON"<br />

Features: *RST value: OFF<br />

SCPI: device-specific<br />

Mode:<br />

R<br />