ESIB<br />

Analyzer - Sweep Control<br />

Sweep Blanking – Gap Sweep<br />

For measurements in the time domain, the GAP SWEEP function offers a high degree of flexibility with<br />

regard to the display of measurement data. With the PRE TRIGGER softkey, it is possible to display<br />

measurements taken before the trigger time. With the GAP TIME softkey, the measurements within a<br />

predefined time range can be blanked. Thus, it is possible to display the rising and falling edge of a<br />

signal with high resolution on a single diagram.<br />

Displayed picture<br />

Trigger<br />

(not displayed)<br />

Displayed picture<br />

Pre-Trigger<br />

Time<br />

Trigger to Gap<br />

Time<br />

Gap<br />

Time<br />

Fig 4-17<br />

Trace blanking in sweep mode.– Gap Sweep<br />

Ref Lev<br />

-10.0 dBm<br />

RBW 1 MHz<br />

VBW 300 kHz<br />

SWT 800 µs<br />

RF Att 20 dB<br />

Unit<br />

(dBm)<br />

Span 0 Hz 80 µs / Div Center 914 MHz<br />

Fig. 4-18<br />

Display of a burst without gaps<br />