Sweep Control - Analyzer<br />

ESIB<br />

Operation steps on the ESIB:<br />

[PRESET]<br />

[MODE]<br />

[↑]<br />

ANALYZER<br />

[CENTER: {802} MHz]<br />

[SPAN<br />

{3.6} MHz]<br />

[REF LVL: {0} dBm: RF ATTEN MANUAL: {10} dB]<br />

[COUPLING: RES BW MANUAL: {30} kHz]<br />

[TRACE 1: DETECTOR: RMS]<br />

[SWEEP:<br />

SWEEPTIME MANUAL: {50} ms;<br />

GATE ON<br />

GATE SETTINGS: GATE MODE EDGE: GATE POL POS: GATE RF POWER<br />

GATE ADJUST: SWEEPTIME MANUAL {1} ms: GATE DELAY {300} µs:<br />

GATE LENGTH: {250} µs]<br />

Note: [KEY] Menu called by the KEY. All indications inside the bracket refer to this menu.<br />

{Number} Numeric value to be entered for the corresponding parameter<br />

SOFTKEY Softkey used to select a parameter or enter a value.<br />

The following figure shows the screen display for gate parameter setting: The vertical lines for gate<br />

delay (GL) and gate length (GL) can be adapted to the burst signal by entering numbers or with the<br />

rollkey.<br />

Fig. 4-16<br />

Setting the GATE DELAY and GATE LENGTH times in the time domain using the GD<br />

and GL lines<br />

On quitting the menu GATE ADJUST, the ESIB returns to spectral representation.<br />

1088.7531.12 4.258 E-15