Operating Manual EMI TEST RECEIVER ESIB7 - Rohde & Schwarz

Operating Manual EMI TEST RECEIVER ESIB7 - Rohde & Schwarz

Operating Manual EMI TEST RECEIVER ESIB7 - Rohde & Schwarz

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

ESIB<br />

Analyzer - Level Display/RF Input<br />

Level Display and RF Input Configuration – LEVEL Key Group<br />

The REF and RANGE keys are used to set the reference level (= maximum RF input level, = overload<br />

limit), the maximum level (= upper grid level) , and the display range of the active window. The INPUT<br />

key sets the characteristics of the RF input (input impedance, selection of input and input attenuation).<br />

Reference Level – REF key<br />

In addition to the reference level (maximum RF input level), the ESIB provides a function to define a<br />

maximum level (uppermost grid line on the display):<br />

For a spectrum analyzer, the upper boundary of the measurement diagram (maximum level) is also the<br />

limit of its dynamic range (reference level). This means, that a signal which goes beyond the upper grid<br />

line also overdrives the analyzer.<br />

For compensation of frequency or antenna characteristics, transducer factors can be applied to increase<br />

the signal level. In this case, the calculated values can be above the reference level without actually<br />

being physically applied to the instrument. In order to display these signals on the grid, an upper grid<br />

level which is different from the reference level of the analyzer can be entered.<br />

LEVEL REF menu:<br />

LEVEL<br />

REF/<br />

UNIT<br />

RANGE<br />

REF LEVEL<br />

REF LEVEL<br />

REF LEVEL<br />

OFFSET<br />

GRID<br />

ABS / REL<br />

REF LEVEL<br />

MAX LEVEL<br />

AUTO<br />

MAX LEVEL<br />

MANUAL<br />

The REF key opens a menu to set the<br />

reference level and the input attenuation<br />

for the active measurement window .<br />

The REF LEVEL softkey is automatically<br />

active with the menu called up and<br />

opens the entry window for manual input<br />

of the reference level.<br />

UNIT<br />

At the same time various settings<br />

concerning the level and the attenuation<br />

display can be made.<br />

RF ATTEN<br />

MANUAL<br />

ATTEN AUTO<br />

NORMAL<br />

The functions for setting the attenuation<br />

are identical to those under the INPUT<br />

key and are described in Section<br />

"Configuration of the RF Input - INPUT<br />

Key".<br />

ATTEN AUTO<br />

LOW NOISE<br />

ATTEN AUTO<br />

LOW DIST<br />

MIXER<br />

LEVEL<br />

1088.7531.12 4.165 E-15