14.06.2014 Views

Operating Manual EMI TEST RECEIVER ESIB7 - Rohde & Schwarz

Operating Manual EMI TEST RECEIVER ESIB7 - Rohde & Schwarz

Operating Manual EMI TEST RECEIVER ESIB7 - Rohde & Schwarz

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

ESIB<br />

SENSe Subsystem<br />

:[SENSe:]CORRection:TRANsducer:DELete<br />

This command deletes the selected transducer factor.<br />

Example: ":CORR:TRAN:DEL"<br />

Features: *RST value: -<br />

SCPI: device-specific<br />

Modes:<br />

R, A<br />

This command is an event and therefore has no *RST value.<br />

Prior to this command, the command SENS:CORR:TRAN:SEL must be sent.<br />

:[SENSe:]CORRection:TSET:CATalog?<br />

This command polls the names of all transducer factors stored on the harddisk.<br />

Syntax of output format:<br />

,,<br />

,,,,,,....,,<br />

<br />

Example: ":CORR:TSET:CAT?"<br />

Mode:<br />

R, A<br />

Feature: *RST value: -<br />

SCPI: device-specific<br />

:[SENSe:]CORRection:TSET:ACTive?<br />

This command reveals the active transducer set. If no transducer set is switched on, a void string will<br />

be output.<br />

Example: ":CORR:TSET:ACT?"<br />

Feature: *RST value: -<br />

SCPI: device-specific<br />

Mode:<br />

R, A<br />

:[SENSe:]CORRection:TSET:SELect <br />

This command selected the transducer set designated by . If does not exist yet, a<br />

new set is created.<br />

Parameter: ::= name of the transducer set in string data form with a maximum<br />

of 8 characters.<br />

Example: ":CORR:TSET:SEL ’SET1’"<br />

Features: *RST value: -<br />

SCPI: device-specific<br />

Modes:<br />

R, A<br />

This command must be sent prior to the subsequent commands for changing/activating the<br />

transducer sets.<br />

1088.7531.12 6.113 E-15

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!