Configuration/Setup<br />

ESIB<br />

Preamplification and Preselection<br />

In the frequency range up to 7 GHz, ESIB has a preselection with switchable preamplifier which can be<br />

selected by the user in the analyzer mode. The preselection is always active in the receiver mode.<br />

Option ESIB-B2, Preamplification, extends the frequency range for preamplification to 26.5 GHz or<br />

40 GHz.<br />

The 20 dB preamplifier is available only if the preselection is switched on.<br />

Preselection<br />

The frequency range 20 Hz to 7 GHz is distributed over nine filter bands. Two fixed-tuned filters are<br />

used up to 2.025 MHz, six tracking passband filters from 2.025 to 1000 MHz and a fixed-tuned highpass<br />

filter above 1 GHz.<br />

The filters are switched by a relay at 150 kHz and by PIN-diode switches above 150 kHz.<br />

LP 150 kHz<br />

BP 150 kHz...2.025MHz<br />

BP 2.025...8.025MHz<br />

Preamplifier<br />

BP 8.025...25.025MHz<br />

20 Hz...7 GHz 20 Hz...7 GHz<br />

BP 25.025...80.025MHz<br />

BP 80.025...200.025MHz<br />

BP 200.025...500.025MHz<br />

BP 500.025 ...1000MHz<br />

HP 1GHz<br />

Fig. 4-6 Preselection and preamplifier<br />

CONFIGURATION SETUP menu:<br />

PRESELECT<br />

ON OFF<br />

The PRESELECT ON/OFF softkey switches the preselection on or off.<br />

The softkey is available only in the analyzer mode.<br />

IEC/IEEE bus command :INPut:PRESelection[:STATe] ON|OFF<br />