ESIB<br />

SENSe Subsystem<br />

:[SENSe:]FREQuency:SPAN:LINK CENTer | START | STOP<br />

This command defines the coupling for frequency-span changes.<br />

Example: ":FREQ:SPAN:LINK STOP"<br />

Features: *RST value: CENTer<br />

SCPI: conforming<br />

Mode:<br />

A<br />

:[SENSe:]FREQuency:STARt 0 GHz to f max<br />

This command defines the start frequency of the scan in receiver mode or the start frequency of the<br />

analyzer.<br />

Example: ":FREQ:STAR 20MHz"<br />

Features: *RST value: 0<br />

SCPI: conforming<br />

Modes:<br />

R, A<br />

In analyzer mode, the automatic coupling of the parameters is set to STOP FIXED.<br />

:[SENSe:]FREQuency:STARt:LINK CENTer | STOP | SPAN<br />

This command defines the coupling for start-frequency changes.<br />

Example: ":FREQ:STAR:LINK SPAN"<br />

Features: *RST value: STOP<br />

SCPI: device-specific<br />

Mode:<br />

A<br />

:[SENSe:]FREQuency:STOP 0 GHz to f max<br />

This command defines the stop frequency of the scan in receiver mode or the stop frequency of the<br />

analyzer.<br />

Example: ":FREQ:STOP 2000MHz"<br />

Features: *RST value: f max<br />

SCPI: conforming<br />

Modes:<br />

R, A<br />

In analyzer mode, the automatic coupling of the parameters is set to STARt FIXED.<br />

:[SENSe:]FREQuency:STOP:LINK CENTer | STARt | SPAN<br />

This command defines the coupling for stop-frequency changes.<br />

Example: ":FREQ:STOP:LINK SPAN"<br />

Features: *RST value: STARt<br />

SCPI: device-specific<br />

Mode:<br />

A<br />