14.06.2014 Views

Operating Manual EMI TEST RECEIVER ESIB7 - Rohde & Schwarz

Operating Manual EMI TEST RECEIVER ESIB7 - Rohde & Schwarz

Operating Manual EMI TEST RECEIVER ESIB7 - Rohde & Schwarz

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

ESIB<br />

List of Error Messages<br />

9 List of Error Messages<br />

The following list contains the error messages for errors occurring in the instrument. The meaning of<br />

negative error codes is defined in SCPI, positive error codes mark errors specific of the instrument.<br />

Error messages are entered in the error/event queue of the status reporting system in the remote<br />

control mode and can be queried with the command SYSTem:ERRor?. The answer format of ESIB to<br />

the command is as follows:<br />

, "; "<br />

The indication of the remote control command with prefixed semicolon is optional.<br />

Example:<br />

The command "<strong>TEST</strong>:COMMAND" generates the following answer to the query SYSTem:ERRor? :<br />

-113,"Undefined header;<strong>TEST</strong>:COMMAND"<br />

The table contains the error code in the left-hand column. In the right-hand column the error text being<br />

entered into the error/event queue or being displayed is printed in bold face. Below the error text, there<br />

is an explanation as to the respective error.<br />

SCPI-Specific Error Messages<br />

No Error<br />

Error code<br />

Error text in the case of queue poll<br />

Error explanation<br />

0 No error<br />

This message is output if the error queue does not contain any entries.<br />

Command Error - Faulty command; sets bit 5 in the ESR register.<br />

Error code<br />

Error text in the case of queue poll<br />

Error explanation<br />

-100 Command Error<br />

The command is faulty or invalid.<br />

-101 Invalid Character<br />

The command contains an invalid sign.<br />

Example: A header contains an ampersand, "SENSe&".<br />

-102 Syntax error<br />

The command is invalid.<br />

Example: The command contains block data the instrument does not accept.<br />

-103 Invalid separator<br />

The command contains an impermissible sign instead of a separator.<br />

Example: A semicolon is missing after the command.<br />

1088.7531.12 9.1 E-2