14.06.2014 Views

Operating Manual EMI TEST RECEIVER ESIB7 - Rohde & Schwarz

Operating Manual EMI TEST RECEIVER ESIB7 - Rohde & Schwarz

Operating Manual EMI TEST RECEIVER ESIB7 - Rohde & Schwarz

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

ESIB- Receiver<br />

ESIB<br />

0 DB MIN<br />

ON OFF<br />

The 0 DB MIN softkey determines whether the 0 dB setting of<br />

the attenuator may be used in the manual or automatic setting<br />

of the attenuator.<br />

0 DB MIN OFF is the default value, ie an RF attenuation of at<br />

least 10 dB is always set on ESIB to protect the input mixer.<br />

An attenuation of 0 dB cannot be set manually either. This<br />

avoids 0 dB being switched on inadvertently particularly when<br />

DUTs with high RFI voltage are measured.<br />

Caution:<br />

If 0 dB RF attenuation is used with<br />

autoranging, care must be taken that the<br />

permissible signal level at the RF input is not<br />

exceeded. Exceeding this level would<br />

damage the input mixer. The 0 dB<br />

attenuation should under no circumstances<br />

be used when RFI voltage measurements<br />

are performed with artificial networks as in<br />

this case very high pulses occur during<br />

phase switching.<br />

C/IEEE-bus command<br />

:INPut:ATTenuation:PROTection ON | OFF<br />

AUTO RANGE<br />

ON OFF<br />

The AUTO RANGE ON/OFF softkey switches the autorange<br />

function on and off.<br />

With the autorange function active, ESIB automatically sets the<br />

attenuation so that a good S/N ratio is obtained without the<br />

receiver stages being overdriven.<br />

IEC/IEEE-bus command<br />

:INPut:ATTenuation:AUTO ON | OFF<br />

AUTOPREAMP<br />

ON OFF<br />

The AUTOPREAMP ON/OFF softkey switches the autopreamp<br />

function and or off.<br />

With the autopreamp function active, the preamplifier is<br />

considered in the autorange procedure. The preamplifier is cut<br />

in when the RF attenuation is reduced to the minimum settable<br />

value.<br />

IEC/IEEE-bus command :INPut:GAIN:AUTO ON | OFF<br />

1088.7531.12 4.78 E-15