ESIB<br />

Receiver - Traces<br />

Trace Selection and Setup –TRACE Key Group<br />

The ESIB can display up to four separate traces at the same time. A trace has a maximum of 500 pixels<br />

along the horizontal frequency axis. If there are more measurement values than pixels available, then<br />

several measurement values are assigned to one pixel.<br />

The traces are selected using keys 1 to 4 of the TRACES key group.<br />

The traces can be individually enabled for a measurement and, after the measurement is completed, be<br />

frozen. Traces which are not enabled remain blanked.<br />

A display mode ican be selected for each trace. The traces can be overwritten (CLEAR/WRITE mode)<br />

after each sweep, or the maximum/ minimum over several sweeps can be displayed.<br />

Individual detectors are selectable for the various traces. The peak detector displays the maximum<br />

values represented by a pixel. The quasi peak detector displays the value of the level at a pixel weighted<br />

according to CISPR. The rms detector displays the power (rms value) of the measured values<br />

represented by a pixel, the average detector the linear average value.<br />

The maximum number of measured frequencies which can be measured is limited and depends on the<br />

number of the traces which are switched on.<br />

Number of traces<br />

Measured values /trace<br />

1 250.000<br />

2 150.000<br />

3 100.000<br />

4 80.000<br />

They are stored for postprocessing. If the scan subranges are defined so that more than the possible<br />

values would be measured, a respective message is output upon the scan start. Afterwards the scan is<br />

performed up to the maximum value.<br />

Measurement Function Selection - TRACE 1 to 4 key<br />

The trace functions are categorized as follows:<br />

• type of trace display (CLEAR/WRITE, VIEW and BLANK)<br />

• evaluation of the trace as a whole (MAX HOLD and MIN HOLD)<br />

• detector of a trace (PEAK, QUASI PEAK, RMS, AVERAGE and AC VIDEO<br />

(AC VIDEO with option ESIB-B1 only)<br />

1088.7531.12 4.143 E-15