CALCulate Subsystem<br />

ESIB<br />

:CALCulate:LIMit:UPPer:SPACing LINear | LOGarithmic<br />

This command makes a selection between linear and logarithmic interpolation for the upper limit line.<br />

Example: ":CALC:LIM:UPP:SPAC LIN"<br />

Features: *RST value: LIN<br />

SCPI: device-specific<br />

Modes:<br />

R, A, VA<br />

:CALCulate:LIMit:LOWer[:DATA] ,..<br />

This command defines the values for the selected lower limit line.<br />

Example: ":CALC:LIM:LOW -30,-40,-40,-30"<br />

Features: *RST value: - (LIMit:STATe is set to OFF)<br />

SCPI: conforming<br />

Modes:<br />

R, A, VA<br />

The number of values for the CONTrol axis and the corresponding LOWer limit line have to be<br />

identical.If the measured values violate the LOWer limit line in analyzer or vector analyzer mode, the<br />

limit check signals errors. The unit must be identical with the unit selected by CALC:LIM:UNIT.<br />

The units DEG, RAD, S, and HZ are available in the vector signal analysis mode only.<br />

:CALCulate:LIMit:LOWer:STATe ON | OFF<br />

This command defines the selected limit line as lower limit line.<br />

Example: ":CALC:LIM:LOWer:STAT ON"<br />

Features: *RST value: OFF<br />

SCPI: conforming<br />

Modes:<br />

R, A, VA<br />

In analyzer and vector analyzer mode, the result of the limit check can be queried with<br />

CALCulate:LIMit:FAIL?.<br />

:CALCulate:LIMit:LOWer:OFFSet <br />

This command defines an offset for the Y-axis of the selected relative lower limit line.<br />

Example: ":CALC:LIM:LOW:OFFS 3dB"<br />

Features: *RST value: 0<br />

SCPI: device-specific<br />

Modes:<br />

A, VA<br />

:CALCulate:LIMit:LOWer:MARGin <br />

This command defines the margin of the selected lower limit line.<br />

Example: ":CALC:LIM:LOW:MARG 10dB"<br />

Features: *RST value: 0<br />

SCPI: device-specific<br />

Modes:<br />

A, VA<br />

1088.7531.12 6.26 E-15