ESIB- Receiver<br />

ESIB<br />

Setting the Measurement Time<br />

The measurement time is the time during which ESIB measures the input signal and forms a<br />

measurement result weighted by the selected detector. The measurement time does not include settling<br />

times. ESIB automatically waits until transients are over.<br />

CONFIGURATION MODE - <strong>EMI</strong> <strong>RECEIVER</strong> menu<br />

MEAS TIME<br />

The MEAS TIME softkey activates the entry field for the measurement time.<br />

The measurement time can be set in steps of 1-2-5 in the range 100 µs to<br />

100 s. A measurement time of 15 s is available in addition.<br />

When the quasi-peak detector is used, the minimum measurement time is<br />

1 ms.<br />

With the average, RMS, AC video or min/max peak detector the smallest<br />

settable measurement time depends on the bandwidth.<br />

Bandwidth Shortest measurement time Shortest measurement time<br />

AV, RMS<br />

PK+, PK-, AC video<br />

≤ 10 Hz 1 s 10 ms<br />

100 Hz 100 ms 1 ms<br />

200 Hz 50 ms 1 ms<br />

1 kHz 10 ms 0.1 ms<br />

9 kHz 1 ms 0.1 ms<br />

≥ 100 kHz 0.1 ms 0.1 ms<br />

IEC/IEEE-bus command<br />

:[SENSe:]SWEep:TIME <br />

Effect of measurement time with different weighting modes:<br />

MIN/MAX PEAK measurement:<br />

With peak weighting selected, the maximum or minimum level<br />

during the selected measurement time is displayed. The peak<br />

detector is reset at the beginning of each measurement. At the<br />

end of the measurement time the maximum or minimum level<br />

occurred during the measurement time is displayed. Since the<br />

peak detector of ESIB is a digital detector, discharging is<br />

irrelevant even with long measurement times.<br />

Unmodulated signals can be measured with the shortest possible<br />

measurement time. When pulses are measured, the selected<br />

measurement time must be long enough for at least one pulse to<br />

occur during the measurement time.<br />

AC video measurement:<br />

Since the AC video detector is a combination of a min peak and a<br />

max peak detector, the considerations for the peak measurement<br />

apply here too<br />

The longer the measurement time, the better the acquisition of<br />

the AC voltage component of the test signal.<br />

1088.7531.12 4.84 E-15