ESIB<br />

Analyzer - Level Display/RF Input<br />

Display Units<br />

Generally, a spectrum analyzer measures the signal voltage at the RF input. The level display is<br />

calibrated in terms of therms values of an unmodulated sine wave. In the initial state, the level is<br />

displayed at a power of 1 milliwatt (= dBm). As the input impedance is known to be 50Ω or 75Ω,<br />

conversion to other units is possible. The units dBm, dBµV, dBµA, dBpW, V, A and W are directly<br />

convertible and can be selected in the REF UNIT menu.<br />

Units dB../MHz have a special status. They are applicable for broadband pulse signals. The measured<br />

pulse voltage or the pulse current is referred to 1 MHz bandwidth. This conversion is not useful for<br />

narrowband or sinewave signals.<br />

If the antenna coded connector on the front panel is used, the coded unit there determines the possible<br />

display units. Connecting to the coded connector deactivates the settings of the UNIT menu.<br />

By certain codings it is nevertheless possible to select a conversion of the unit in the menu. The<br />

dependencies between the unit of the antenna coded connector or the unit of the transducer and the<br />

unit which is to be selected for the display are given in the softkey description.<br />

The default coding defined by the connector can be switched off using the PROBE CODE ON/OFF<br />

softkey. In this case, the unit can be set with the corresponding unit softkeys (dBm, dBµV, ... ) even if a<br />

coded connector is plugged in. The coded entries in the connector are then ignored.<br />

Note:<br />

The units dBµV/m and dBµA/m can only be set via the unit of a transducer or via the unit of the<br />

coded connector.<br />

LEVEL REF UNIT submenu:<br />

UNIT<br />

UNIT<br />

dBm<br />

dBmV<br />

dBµV<br />

dBµA<br />

The UNIT softkey opens a sub-menu in which the desired<br />

units for the level axis can be selected and the coding of the<br />

antenna coded connector can be switched on and off.<br />

The selected units are valid for both displays when two<br />

measurement windows are displayed.<br />

The unit softkeys are mutually exclusive selection switches.<br />

Only one of these switches may be active at any one time.<br />

dBpW<br />

dB* / MHz<br />

VOLT<br />

AMPERE<br />

WATT<br />

PROBE CODE<br />

ON / OFF<br />

1088.7531.12 4.167 E-15