Configuration/Setup<br />

ESIB<br />

DELETE<br />

FACTOR/SET<br />

The DELETE FACTOR/SET softkey deletes the marked factor or set. To<br />

avoid deletion by mistake, deletion has to be confirmed.<br />

MESSAGE<br />

Do you really want to<br />

delete factor or set?<br />

YES<br />

NO<br />

IEC/IEEE-bus command<br />

:[SENSe:]CORRection:TRANsducer:DELete<br />

:[SENSe:]CORRection:TSET:DELete<br />

PAGE UP<br />

The PAGE UP and PAGE DOWN softkeys are used to scroll in large tables<br />

which cannot fully be displayed on the screen.<br />

PAGE DOWN<br />

New Entry and Editing of Transducer Factors<br />

A transducer factor is characterized by<br />

• reference values with frequency and transducer factor (Values)<br />

• the unit of the transducer factor (Unit) and<br />

• the name (Name) to distinguish between the different factors.<br />

During the entry, ESIB checks the transducer factor according to specific rules that have to be met to<br />

ensure trouble-free operation.<br />

• The frequencies for the reference values always have to be entered in ascending order. Otherwise,<br />

the entry is not accepted and the following message is displayed:<br />

Frequency Sequence!<br />

• The entered frequencies do not necessarily have to be set at ESIB, as only the values for the<br />

frequency display range are considered for a set sweep or scan. The minimum frequency for a<br />

reference value is 0 Hz, the maximum frequency 200 GHz.<br />

• The minimum or maximum value for a transducer factor is -200 dB or 200 dB. The unit "dB" means<br />

that the transducer factor is always logarithmic and has nothing to do with the physical transducer<br />

factor, which, for example, establishes the relationship between field strength and voltage into 50<br />

Ohm. If the minimum or maximum value is exceeded, ESIB signals:<br />

Min Level -200 dB or<br />

Max Level 200 dB.<br />

• Amplifiers have a negative transducer factor, attenuation values have to be entered as a positive<br />

transducer factor.<br />

Note:<br />

A unit activated by switching on the transducer has priority over a coded unit of the<br />

connected probe.<br />

With the exception of dB*/MHz, the softkeys for the unit in the menu under key LEVEL REF<br />

are inoperative with the transducer switched on.<br />

1088.7531.12 4.24 E-15