STATus Subsystem<br />

ESIB<br />

:STATus:QUEStionable:TRANsducer:ENABle 0 to 65535<br />

This command sets the bits of the ENABle section of the :STATus:QUEStionable:TRANsducer<br />

register.<br />

Example: ":STAT:QUES:TRAN:ENAB 65535"<br />

Features: *RST value: –<br />

SCPI: device-specific<br />

Modes:<br />

R, A, VA<br />

The ENABle register selectively enables the individual events of the associated EVENt section for the<br />

sum bit in the status byte.<br />

:STATus:QUEStionable:TRANsducer:PTRansition 0 to 65535<br />

This command sets the edge detectors of all bits of the :STATus:QUEStionable:TRANsducer<br />

register from 0 to 1 for the transitions of the CONDition bit.<br />

Example: ":STAT:QUES:TRAN:PTR 65535"<br />

Features: *RST value: –<br />

SCPI: device-specific<br />

Modes:<br />

R, A, VA<br />

:STATus:QUEStionable:TRANsducer:NTRansition 0 to 65535<br />

This command sets the edge detectors of all bits of the :STATus:QUEStionable:TRANsducer<br />

register from 1 to 0 for the transitions of the CONDition bit.<br />

Example: ":STAT:QUES:TRAN:NTR 65535"<br />

Features: *RST value: –<br />

SCPI: device-specific<br />

Modes:<br />

R, A, VA<br />

:STATus:QUEue[:NEXT]?<br />

This command queries the earliest entry to the error queue, thus deleting it.<br />

Example: ":STAT:QUE?"<br />

Features: *RST value: –<br />

SCPI: conforming<br />

Modes:<br />

R, A, VA<br />

Positive error numbers indicate device-specific errors, negative error numbers are error messages<br />

defined by SCPI (cf. Chapter 9). If the error queue is empty, the error number 0, "no error", is<br />

returned. This command is identical with the command SYSTem:ERRor.<br />