14.06.2014 Views

Operating Manual EMI TEST RECEIVER ESIB7 - Rohde & Schwarz

Operating Manual EMI TEST RECEIVER ESIB7 - Rohde & Schwarz

Operating Manual EMI TEST RECEIVER ESIB7 - Rohde & Schwarz

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

ESIB<br />

Analyzer - Sweep Control<br />

SWEEP SWEEP-GAP SWEEP SETTINGS sub-menu:<br />

GAP SWEEP<br />

SETTINGS<br />

GAP SWEEP<br />

SETTINGS<br />

TRIGGER<br />

LEVEL<br />

PRE<br />

TRIGGER<br />

TRG TO GAP<br />

TIME<br />

GAP<br />

LENGTH<br />

The GAP SWEEP SETTINGS softkey opens a sub-menu in<br />

which the parameters for the gap can be selected.<br />

The trigger time corresponds to t = 0. Events occurring<br />

before the trigger time are displayed in negative time.<br />

TRIGGER<br />

LEVEL<br />

The TRIGGER LEVEL softkey activates the entry window for trigger level.<br />

This function corresponds to the setting in the trigger menu.<br />

IEC/IEEE-bus command<br />

:TRIGger[:SEQuence]:LEVel:VIDeo 50PCT<br />

PRE<br />

TRIGGER<br />

The PRE TRIGGER softkey activates the entry window for the pre-trigger<br />

time. The pre-trigger time defines the separation between the left edge of the<br />

grid and the trigger time (t = 0).<br />

At the same time the gap sweep is switched on (exception : entry t = 0)<br />

The minimum pre-trigger time is −100 s, the maximum pre-trigger time<br />

depends on the sweep time and the trigger to gap time (max = 100 s). The<br />

maximum resolution is 50 ns.<br />

The PRE TRIGGER value can be entered in the frequency domain (span > 0)<br />

as well as in the time domain and GAP SWEEP OFF mode. It is, however,<br />

only effective when the GAP SWEEP measurement is enabled.<br />

IEC/IEEE-bus command<br />

:SENSe:]SWEep:GAP:PRETrigger 100us<br />