Messages<br />

ESIB<br />

Messages<br />

The messages transferred via the data lines of the IEC bus or the RSIB interface (see chapter 8) can be<br />

divided into two groups:<br />

– interface messages and<br />

– device messages.<br />

Some control characters are defined for the control of the RS-232-interface (see chapter 8).<br />

IEE/IEEE-Bus Interface Messages<br />

Interface messages are transferred on the data lines of the IEC bus, the "ATN" control line being active.<br />

They are used for communication between controller and instrument and can only be sent by a<br />

controller which has the IEC-bus control. Interface commands can be subdivided into<br />

– universal commands and<br />

– addressed commands.<br />

Universal commands act on all devices connected to the IEC bus without previous addressing,<br />

addressed commands only act on devices previously addressed as listeners. The interface messages<br />

relevant to the instrument are listed in chapter 8.<br />

RSIB Interface Messages<br />

The RSIB interface enables the instrument to be controlled by Visual C++ or Visual Basic programs. The<br />

interface functions are matched to the function interface for IEC/IEEE-bus programming from National<br />

Instruments.<br />

The functions supported by interface are listed in chapter 8.<br />