Status Reporting System<br />

ESIB<br />

STATus QUEStionable:TRANsducer Register<br />

This register indicates that a transducer hold point is attained (bit 15) and what range is to be swept next<br />

(bit 0 to 10). The sweep can be continued with command INITiate2:CONMeasure.<br />

It can be queried with commands STATus:QUEStionable:TRANsducer:CONDition? and "STATus<br />

:QUEStionable:TRANsducer[:EVENt]?.<br />

Table 5-11<br />

Meaning of bits in STATus:QUEStionable:TRANsducer register<br />

Bit No.<br />

Meaning<br />

0<br />

1<br />

Range 1<br />

This bit is set when subrange 1 is attained.<br />

Range 2<br />

This bit is set when subrange 2 is attained.<br />

2 Range 3<br />

This bit is set when subrange 3 is attained.<br />

3 Range 4<br />

This bit is set when subrange 4 is attained.<br />

4 Range 5<br />

This bit is set when subrange 1 is attained.<br />

5 Range 6<br />

This bit is set when subrange 6 is attained.<br />

6 Range 7<br />

This bit is set when subrange 7 is attained.<br />

7 Range 8<br />

This bit is set when subrange 8 is attained.<br />

8 Range 9<br />

This bit is set when subrange 9 is attained.<br />

9 Range 10<br />

This bit is set when subrange 10 is attained.<br />

10 not used<br />

11 not used<br />

12 not used<br />

13 not used<br />

14 Subrange limit<br />

This bit is set when the transducer is at the point of changeover from one range to another.<br />

15 This bit is always 0.<br />

1088.7531.12 5.30 E-16