SENSe Subsystem<br />

ESIB<br />

:[SENSe:]MIXer:PORTs 2 | 3<br />

This command activates the 2- or 3-port mixer. In the BAND LOCK ON mode, the command refers<br />

to the active band selected with SENSe:MIXer:HARMonic:BAND .<br />

Example: ":MIX:PORT 3"<br />

Features: *RST value: 2<br />

SCPI: device-specific<br />

Mode:<br />

A<br />

This command is available only if the external mixer (option FSE-B21) is switched on.<br />

:[SENSe:]MIXer:SIGNal ON | OFF | AUTO<br />

This command activates the Signal ID or Auto ID mode.<br />

Example: ":MIX:SIGN ON"<br />

Features: *RST value: OFF<br />

SCPI: device-specific<br />

Mode:<br />

A<br />

This command is available only if the external mixer (option FSE-B21) is switched on.<br />

:[SENSe:]MIXer:THReshold 0.1 to 100 dB<br />

This command sets the level threshold for auto ID.<br />

Example: ":MIX:THR 20"<br />

Features: *RST value: 10<br />

SCPI: device-specific<br />

Mode:<br />

A<br />

The command is available only if the external mixer (option FSE-B21) is switched on.<br />

:[SENSe:]MIXer:HARMonic 1 to 62<br />

With BAND LOCK OFF, this command sets the nth harmonic. The command may be a query with<br />

BAND LOCK ON.<br />

Example: ":MIX:HARM 5"<br />

Features: *RST value: 2<br />

SCPI: conforming<br />

Mode:<br />

A<br />

The command is available only if the external mixer (option FSE-B21) is switched on.<br />

:[SENSe:]MIXer:HARMonic:TYPE ODD | EVEN | EODD<br />

With BAND LOCK ON, this command sets the type of harmonic.<br />

Example: ":MIX:HARM:TYPE EODD"<br />

Features: *RST value: EVEN<br />

SCPI: device-specific<br />

Mode:<br />

A<br />

The command is available only if the external mixer (option FSE-B21) is switched on.<br />

1088.7531.12 6.138 E-15