ESIB<br />

Setting the Step Size<br />

Setting the Stepsize - STEP Key<br />

A number of numeric instrument parameters allow for step-by-step incrementing or decrementing the<br />

value in the editing line of the entry window using the keys or or the roll-key. The stepsize used<br />

with the cursor keys can be specified for the selected parameters in the STEP menu.<br />

Notes: – It is not possible to set the stepsize by all of the numeric parameters.<br />

– An extended STEP menu is available by some parameters. The extra functions are<br />

described by the respective parameter.<br />

– Setting the step size has no effect on the roll-key (the resolution of the roll-key is higher than<br />

that of the step size functuion)<br />

D A TA V AR I A T I ON<br />

The STEP key opens the STEP menu for entry of the stepsize.<br />

STEP<br />

The stepsize of an instrument parameter can only be varied when this<br />

parameter is edited in an entry window.<br />

The softkeys in the STEP menu are disabled if no instrumentnparameter is<br />

being varied or if the parameter being edited does not support stepsize<br />

variation (as is the case by all of the alphanumeric parameters.<br />

The key leads to exit from the STEP menu and automatic return to the<br />

previously active menu.<br />

STEPSIZE<br />

AUTO<br />

The softkey STEPSIZE AUTO causes the stepsize for the corresponding<br />

instrument parameter to be set automatically and be continuously varied with<br />

varying instrument settings (in accordance with other parameters).<br />

Example:<br />

The stepsize of the CENTER frequency depends on the selected span.<br />

STEPSIZE<br />

MANUAL<br />

The softkey STEPSIZE NORMAL calls an entry window for definition of a<br />

stepsize by the user. The parameter whose stepsize is varied, is displayed in<br />

the headline of the entry window:<br />

START FREQUENCY STEPSIZE<br />

100 kHz<br />

The stepsize remains constant until a new value is entered or until the<br />

STEPSIZE AUTO softkey is activated. It is no longer coupled to other<br />

instrument parameters.<br />