R&S <strong>FSQ</strong> Instrument Functions<br />

RES BW<br />

MANUAL<br />

VIDEO BW<br />

MANUAL<br />

SWEEPTIME<br />

MANUAL<br />

<strong>Analyzer</strong> Mode<br />

The RES BW MANUAL softkey activates the manual data entry for the resolution<br />

bandwidth.<br />

The resolution bandwidth can be selected in 1/2/3/5/10 steps in the range between<br />

10 Hz and 20 MHz. Additionally a maximum bandwidth of 50 MHz is available. The<br />

nominal resolution bandwidth is the 3 dB bandwidth.<br />

When FFT filters are used, the lower limit of the bandwidth is 1 Hz. FFT filters may<br />

be used with bandwidths up to 30 kHz.<br />

For numeric inputs, the values are always rounded to the nearest possible bandwidth.<br />

For rotary knob or UP/DOWN key inputs, the bandwidth is adjusted in steps<br />

either upwards or downwards.<br />

For filter type CHANNEL or RRC, the bandwidth is selected from the list of available<br />

channel filters given in section “Filter Types” on page 4.26.<br />

For data entry, the cursor keys Uu and Ud scroll through this list.<br />

The manual input mode of the resolution bandwidth is indicated by a green asterisk<br />

(*) on the display.<br />

Remote command: BAND:AUTO OFF;<br />

BAND 1MHz<br />

The VIDEO BW MANUAL softkey activates the manual data entry for the video<br />

bandwidth.<br />

The video bandwidth can be selected in 1/2/3/5/10 steps in the range between 1 Hz<br />

and 10 MHz.<br />

For numeric inputs, the values are always rounded to the nearest possible allowed<br />

bandwidth. For rotary knob or UP/DOWN key inputs, the bandwidth is adjusted in<br />

steps either downwards or upwards.<br />

The manual input mode of the video bandwidth is indicated by a green asterisk (*)<br />

on the display.<br />

Remote command: BAND:VID:AUTO OFF;<br />

BAND:VID 10 kHz<br />

The SWEEPTIME MANUAL softkey activates the manual data entry for the sweep<br />

time. At the same time, the coupling of the sweep time is cancelled. Other couplings<br />

(VIDEO BW, RES BW) remain effective.<br />

In the frequency domain (span > 0 Hz) and for resolution bandwidths above 1 kHz,<br />

the allowed sweep times for spans > 3.2 kHz range from 2.5 ms through to 16000 s.<br />

With spans below 3.2 kHz, the maximum allowed sweep time is reduced to 5 s *<br />

span/Hz.<br />

If FFT filters are used, the sweep time is fixed by the span and the bandwidth and<br />

therefore cannot be set.<br />

In time domain (span = 0 Hz), the range of sweep times is 1 µs to 16000 s is selectable<br />

in steps of max. 5% of the sweep time. For numeric inputs, rounding is made to<br />

the nearest possible sweep time. For rotary knob or UP/DOWN key inputs, the<br />

sweep time is adjusted in steps either downwards or upwards.<br />

4.22 Operating Manual 1313.9681.12 - 01