R&S <strong>FSQ</strong> Instrument Functions<br />

AUTO MAX<br />

PEAK<br />

AUTO MIN<br />

PEAK<br />

Order of signals detected:<br />

PEAK: signal 1<br />

NEXT PEAK: signal 1 (no further signal detected)<br />

or<br />

PEAK: signal 1<br />

NEXT PEAK LEFT: signal 1 (no further signal detected)<br />

NEXT PEAK RIGHT: signal 1 (no further signal detected)<br />

<strong>Analyzer</strong> Mode<br />

With 20 dB peak excursion, NEXT PEAK and NEXT PEAK RIGHT will also detect<br />

signal 2, as the signal level decreases at least by 29.85 dB to either side of this signal,<br />

which is now greater than the peak excursion.<br />

Order of signals detected:<br />

PEAK: <strong>Signal</strong> 1<br />

NEXT PEAK: <strong>Signal</strong> 2<br />

NEXT PEAK: <strong>Signal</strong> 2 (no further signal detected)<br />

or<br />

PEAK: <strong>Signal</strong> 1<br />

NEXT PEAK LEFT: <strong>Signal</strong> 1 (no further signal detected)<br />

NEXT PEAK RIGHT: <strong>Signal</strong> 2<br />

NEXT PEAK RIGHT: <strong>Signal</strong> 2 (no further signal detected)<br />

With 6 dB peak excursion, all signals will be detected with NEXT PEAK and NEXT<br />

PEAK RIGHT / NEXT PEAK LEFT.<br />

Order of signals detected:<br />

PEAK: <strong>Signal</strong> 1<br />

NEXT PEAK: <strong>Signal</strong> 2<br />

NEXT PEAK: <strong>Signal</strong> 3<br />

NEXT PEAK: <strong>Signal</strong> 4<br />

or<br />

PEAK: <strong>Signal</strong> 1<br />

NEXT PEAK LEFT: <strong>Signal</strong> 3<br />

NEXT PEAK RIGHT: <strong>Signal</strong> 1<br />

NEXT PEAK RIGHT: <strong>Signal</strong> 2<br />

NEXT PEAK RIGHT. <strong>Signal</strong> 4<br />

The AUTO MAX PEAK / AUTO MIN PEAK softkeys add an automatic peak search<br />

action for marker 1 at the end of each particular sweep. This function may be used<br />

during adjustments of a device under test to keep track of the current peak marker<br />

position and level.<br />

4.78 Operating Manual 1313.9681.12 - 01