R&S <strong>FSQ</strong> Remote Control – Description of Commands<br />

Mode: A<br />

SOURce:EXTernal:ROSCillator[:SOURce] INTernal | EXTernal<br />

SOURce Subsystem<br />

This command switches between external and internal reference oscillator for the frequency<br />

processing of external generator 1 and 2.<br />

The command always works on both generators. Therefore, the numeric suffix in EXTernal is not<br />

significant.<br />

This command is only valid in combination with option Ext. Generator Control R&S FSP-B10.<br />

Example: "SOUR:EXT:ROSC EXT"<br />

'switches to external reference oscillator<br />

Characteristics: *RST value: INT<br />

SCPI: device-specific<br />

Mode: A<br />

This command is available from firmware version 1.40 or higher.<br />

SOURce:EXTernal[:STATe] ON | OFF<br />

This command activates or deactivates the external generator selected with SOUR:EXT:FREQ:<br />

SWE ON in the selected window.<br />

The suffix behind EXTernal is irrelevant for this command.<br />

This command is only available in connection with option Ext. Generator Control B10.<br />

Example: "SYST:COMM:RDEV:GEN1:TYPE 'SMP02'"<br />

'Selects SMP02 as generator 1.<br />

"SYST:COMM:RDEV:GEN1:LINK TTL"<br />

'Selects GPIB + TTL link as interface.<br />

"SYST:COMM:RDEV:GEN1:ADDR 28"<br />

'Sets the generator address to 28.<br />

"SOUR:EXT1:FREQ:SWE ON"<br />

'Activates the frequency sweep for generator 1.<br />

"SOUR:EXT ON"<br />

'Activates the external generator<br />

Characteristics: *RST value: OFF<br />

SCPI: device-specific<br />

Mode: A<br />