R&S <strong>FSQ</strong> Remote Control – Description of Commands<br />

6.21 TRIGger Subsystem<br />

TRIGger Subsystem<br />

The TRIGger subsystem is used to synchronize instrument actions with events. It is thus possible to control<br />

and synchronize the start of a sweep. An external trigger signal can be applied to the connector at<br />

the rear panel of the instrument. A distinction is made between TRIGger1 (screen A) and TRIGger2<br />

(screen B).<br />

TRIGger[:SEQuence]:HOLDoff -100 to 100s<br />

This command defines the length of the trigger delay.<br />

A negative delay time (pretrigger) can be set in the time domain (span = 0 Hz) only.<br />

Example: "TRIG:HOLD 500us"<br />

Characteristics: *RST value: 0 s<br />

SCPI: conform<br />

Mode: A<br />

TRIGger[:SEQuence]:HOLDoff:ADJust:AUTO ON|OFF<br />

This command defines whether the group delay of the bandwidth filters for the external or IF power<br />

trigger is compensated (ON) or not (OFF). If a burst signal is analyzed in the time domain mode, the<br />

rising edge will remain at the same time position if compensation is on and the bandwidth changes.<br />

Example: "TRIG:HOLD:ADJ:AUTO ON"<br />

Characteristics: *RST value: OFF<br />

SCPI: device-specific<br />

Mode: A<br />

TRIGger[:SEQuence]:LEVel[:EXTernal] 0.5 to +3.5 V<br />

This command sets the level of the external trigger source.<br />

Example: "TRIG:LEV 2V"<br />

Characteristics: *RST value: 1.4 V<br />

SCPI: device-specific<br />

Mode: A<br />

TRIGger[:SEQuence]:LEVel:IFPower -70 to +30 dBm<br />

This command sets the level of the IF power trigger source.<br />

Example: "TRIG:LEV:IFP –20DBM"<br />

Characteristics: *RST value: -20 dBm<br />

SCPI: device-specific<br />

Mode: A<br />

Operating Manual 1313.9681.12 - 01 6.253