18.12.2012 Views

R&S FSQ Signal Analyzer - Rohde & Schwarz

R&S FSQ Signal Analyzer - Rohde & Schwarz

R&S FSQ Signal Analyzer - Rohde & Schwarz

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

R&S <strong>FSQ</strong> Instrument Functions<br />

• Transd. Factor<br />

<strong>Analyzer</strong> Mode<br />

Shows the name of the transducer factor that is used for the measurement, or<br />

NONE if no transducer factor is selected.<br />

To select a transducer factor highlight this field with the rotary knob and press the<br />

ENTER key to open the transducer selection list. Select the name of the<br />

transducer factor you need with the rotary knob or NONE if you have to work<br />

without a transducer factor.<br />

Remote command: ESP:RANG1:TRAN <br />

• Limit Check<br />

Sets the type of limit check for all ranges.<br />

ABSOLUTE:<br />

Checks only the defined absolute limits.<br />

RELATIVE:<br />

Checks only the relative limits. Relative limits are defined as relative to the<br />

measured power in the reference range.<br />

ABS & REL:<br />

Combines the absolute and relative limit. The limit check fails when both limits are<br />

violated.<br />

ABS or REL:<br />

Combines the absolute and relative limit. The limit check fails when one of the<br />

limits is violated.<br />

Remote command: ESP:RANG1:LIM:STAT ABS | REL | AND | OR<br />

• Abs Limit Start<br />

Sets the absolute limit value at the start frequency of the range. The unit depends<br />

on the unit selected via the UNIT softkey.<br />

Remote command: ESP:RANG1:LIM:ABS:STAR 10<br />

• Abs Limit Stop<br />

Sets the absolute limit value at the stop frequency of the range. The unit depends<br />

on the unit selected via the UNIT softkey.<br />

Remote command: ESP:RANG1:LIM:ABS:STOP 20<br />

• Rel Limit Start<br />

Sets the relative limit value at the start frequency of the range. The unit is dBc.<br />

Remote command: ESP:RANG1:LIM:REL:STAR -20<br />

• Rel Limit Stop<br />

Sets the relative limit value at the stop frequency of the range. The unit is dBc.<br />

Remote command: ESP:RANG1:LIM:REL:STOP -20<br />

Operating Manual 1313.9681.12 - 01 4.149