R&S <strong>FSQ</strong> Instrument Functions<br />

NEW<br />

FACTOR /<br />

EDIT TRD FAC<br />

TOR<br />

Basic Settings<br />

• The frequencies entered may exceed the frequency range of the R&S <strong>FSQ</strong> since<br />

only the set frequency range is taken into account for measurements. The<br />

minimum frequency of a reference value is 0 Hz, the maximum frequency<br />

200 GHz.<br />

• The value range for the transducer factor is ±200 dB. If the minimum or maximum<br />

value is exceeded, the R&S <strong>FSQ</strong> outputs the following message:<br />

out of range<br />

• Gain has to be entered as a negative value, and attenuation as a positive value.<br />

The softkeys in the UNIT submenu of the AMPT key cannot be operated if the<br />

transducer is on.<br />

The NEW FACTOR and EDIT TRD FACTOR softkeys give access to the submenu<br />

for editing and generating transducer factors.<br />

TRD FACTOR NAME<br />

TRD FACTOR UNIT<br />

TRD FACTOR VALUES<br />

INSERT LINE<br />

DELETE LINE<br />

SAVE TRD FACTOR<br />

Depending on the softkey selected, either the table with the data of the factor<br />

marked (EDIT TRD FACTOR softkey) or an empty table (NEW FACTOR softkey) is<br />

displayed. This table is empty except for the following entries:<br />

Unit: dB<br />

Interpolation: LIN for linear frequency scaling<br />

LOG for logarithmic frequency scaling<br />

The features of the factor are entered in the header of the table, and the frequency<br />

and the transducer factor are entered in the columns.<br />

Name Entry of name<br />

Unit Selection of unit<br />

Interpolation Selection of interpolation<br />

Comment Entry of comment<br />

FREQUENCY Entry of frequency of reference values<br />

TDF/dB Entry of transducer factor.<br />

4.180 Operating Manual 1313.9681.12 - 01