R&S <strong>FSQ</strong> Instrument Functions<br />

IF GAIN<br />

(NORM PULS)<br />

SWEEP REP<br />

(ON OFF)<br />

COUPLING<br />

(FSP HP)<br />

REV STRING<br />

FACTORY<br />

REV STRING<br />

USER<br />

Basic Settings<br />

The IF GAIN (NORM PULS) softkey configures the internal IF gain settings in HP<br />

emulation mode due to the application needs. This setting is only taken into account<br />

for resolution bandwidth < 300 kHz.<br />

NORM Optimized for high dynamic range,<br />

overload limit is close to reference level.<br />

PULS Optimized for pulsed signals,<br />

overload limit up to 10 dB above reference level.<br />

This softkey is only available, if a HP languge is selected via the GPIB LANGUAGE<br />

softkey.<br />

Remote command: SYST:IFG:MODE PULS<br />

The SWEEP REP (ON OFF) softkey controls a repeated sweep of the E1 and MKPK<br />

HI HP model commands (for details on the commands refer to “GPIB Commands of<br />

HP Models 856xE, 8566A/B, 8568A/B and 8594E” on page 6.256). If the repeated<br />

sweep is OFF, the marker is set without sweeping before.<br />

In single sweep mode, switch off this softkey before you set the marker via the E1<br />

and MKPK HI commands in order to avoid sweeping afresh.<br />

This softkey is only available, if a HP languge is selected via the GPIB LANGUAGE<br />

softkey.<br />

Remote command: SYST:RSW ON<br />

The COUPLING (FSP HP) softkey controls the default coupling ratios for:<br />

• span and resolution bandwidth (Span/RBW) and<br />

• resolution bandwidth and video bandwidth (RBW/VBW)<br />

for the HP emulation mode.<br />

In case of FSP the standard parameter coupling of the instrument is used. As a<br />

result in most cases a shorter sweeptime is used than in case of HP.<br />

The softkey is only available in HP emulation mode.<br />

Remote command: SYST:HPC FSP<br />

The REV STRING FACTORY softkey selects the default response for the “REV?”<br />

remote command. It is available if a HP language is selected (“GPIB LANGUAGE”<br />

softkey).<br />

Remote command: SYST:REV:FACT<br />

The REV STRING USER softkey opens a dialog box to define a customized<br />

response for the “REV?” remote command. The maximum length of the output string<br />

is 40 characters.<br />

The softkey is available if a HP language is selected (“GPIB LANGUAGE” softkey).<br />

Remote command: SYST:REV <br />