R&S <strong>FSQ</strong> Remote Control – Description of Commands<br />

[SENSe:]ESPectrum:RANGe[:FREQuency]:STOP <br />

SENSe Subsystem<br />

This command sets the relative stop frequency of a range in the Spectrum Emission Mask<br />

measurement.<br />

The numeric suffixes specify the range. The numeric suffix at SENSe is irrelevant.<br />

Example: "ESP:RANG2:STOP 2.52MHZ"<br />

'stop frequency of range 2 to +2.52MHz<br />

Characteristics: *RST value:-<br />

SCPI: device-specific<br />

Mode: A<br />

[SENSe:]ESPectrum:RANGe:INPut:ATTenuation <br />

This command defines the RF attenuation of a range in the Spectrum Emission Mask measurement.<br />

The numeric suffixes specify the range. The numeric suffix at SENSe is irrelevant.<br />

Example: "ESP:RANG2:INP:ATT 30db"<br />

'RF attenuation of range 2 to 30 dB<br />

Characteristics: *RST value:-<br />

SCPI: device-specific<br />

Mode: A<br />

[SENSe:]ESPectrum:RANGe:INPut:ATTenuation:AUTO ON | OFF<br />

This command switches the attenuation mode of a range in the Spectrum Emission Mask<br />

measurement to AUTO (ON) or MANUAL (OFF).<br />

The numeric suffixes specify the range. The numeric suffix at SENSe is irrelevant.<br />

Example: "ESP:RANG2:INP:ATT:AUTO ON"<br />

'attenuation mode is AUTO for range 2<br />

Characteristics: *RST value: ON<br />

SCPI: device-specific<br />

Mode: A<br />

[SENSe:]ESPectrum:RANGe:INPut:GAIN:STATe ON | OFF<br />

This command switches the preamplifier of a range in the Spectrum Emission Mask measurement on<br />

or off.<br />

The numeric suffixes specify the range. The numeric suffix at SENSe is irrelevant.<br />

Example: "ESP:RANG2:INP:GAIN:STAT ON"<br />

' activates the preamplifier for range 2<br />

Characteristics: *RST value: OFF<br />

SCPI: device-specific<br />

Mode: A<br />