R&S <strong>FSQ</strong> Remote Control – Description of Commands<br />

[SENSe:]POWer:BANDwidth|BWIDth 10 to 99.9PCT<br />

SENSe Subsystem<br />

This command defines the percentage of the power with respect to the total power. This value is the<br />

basis for the occupied bandwidth measurement (command: POWer:ACHannel:PRESet OBW).<br />

The command is only available in the frequency domain (span > 0).<br />

Example: "POW:BWID 95PCT"<br />

Characteristics: *RST value: 99PCT<br />

SCPI: device-specific<br />

Mode: A-F<br />

[SENSe:]POWer:HSPeed ON | OFF<br />

This command switches on or off the high-speed channel/adjacent channel power measurement. The<br />

measurement itself is performed in the time domain on the center frequencies of the individual<br />

channels. The command automatically switches to the time domain and back.<br />

Depending on the selected mobile radio standard, weighting filters with √cos characteristic or very<br />

steep-sided channel filters are used for band limitation.<br />

The command is only available in the frequency domain (span > 0).<br />

Example: "POW:HSP ON"<br />

Characteristics: *RST value: OFF<br />

SCPI: device-specific<br />

Mode: A-F<br />

[SENSe:]POWer:NCORrection ON | OFF<br />

This command switches on or off the correction of the instrument inherent noise for channel power<br />

measurement. On activating this function, a reference measurement of the instrument inherent noise<br />

is performed. The measured noise power is then subtracted from the power in the examined channel.<br />

The noise correction is avilable for:<br />

• Spectrum analyzer frequency domain measurements<br />

• ACP measurements<br />

• Spectrum emission mask measurements<br />

• Spurious measurements<br />

Example: "POW:NCOR ON"<br />

Characteristics: *RST value: OFF<br />

SCPI: device-specific<br />

Mode: A<br />

If the high-speed power measurement is switched off, the complete channel/adjacent<br />

channel power measurement is switched off.<br />