R&S <strong>FSQ</strong> Instrument Functions<br />

Example<br />

<strong>Analyzer</strong> Mode<br />

A spectrum is displayed with a large span after PRESET. A signal off the center is to<br />

be examined in detail:<br />

[PRESET] R&S <strong>FSQ</strong> is set to the default setting.<br />

[MKR->] MARKER 1 is switched on and automatically jumps to<br />

the largest signal of the trace.<br />

[CENTER=MKR FREQ] The center frequency is set to the marker frequency.<br />

The span is adapted in such a way that the minimum frequency<br />

(= 0 Hz) or the maximum frequency is not<br />

exceeded.<br />

[REF LEVEL = MKR LVL] The reference level is set to the measured marker level.<br />

[SPAN] The span can, for example, be reduced using the rotary<br />

knob.<br />

NEXT PEAK The NEXT PEAK softkey sets the active marker/delta marker to the next lower maximum<br />

of the selected trace.<br />

NEXT PEAK<br />

RIGHT<br />

NEXT PEAK<br />

LEFT<br />

SEARCH<br />

LIMITS<br />

LEFT LIMIT /<br />

RIGHT LIMIT<br />

Remote command: CALC:MARK:MAX:NEXT<br />

CALC:DELT:MAX:NEXT<br />

The NEXT PEAK RIGHT softkey sets the active marker/delta marker to the next<br />

lower maximum right of the current marker position on the selected trace.<br />

Remote command: CALC:MARK:MAX:RIGH<br />

CALC:DELT:MAX:RIGH<br />

The NEXT PEAK LEFT softkey sets the active marker/delta marker to the next lower<br />

maximum left of the current marker position the selected trace.<br />

Remote command: CALC:MARK:MAX:LEFT<br />

CALC:DELT:MAX:LEFT<br />

LEFT LIMIT /<br />

RIGHT LIMIT<br />

THRESHOLD<br />

SEARCH LIMIT OFF<br />

The SEARCH LIMITS softkey limits the search range for maximum or minimum<br />

search. The softkey switches to a submenu in which the search range limits can be<br />

set in the x and y direction.<br />

The LEFT LIMIT and RIGHT LIMIT softkeys define the two vertical lines F1 and F2<br />

in the frequency domain (span > 0) and T1 / T2 in the time domain (span = 0). The<br />

search is performed between these lines in the frequency and time domain<br />

If only LEFT LIMIT is enabled, line F1/T1 is the lower limit and the upper limit corresponds<br />

to the stop frequency. If RIGHT LIMIT is also enabled, it determines the<br />

upper limit.<br />

Remote command: CALC:MARK:X:SLIM:LEFT 1MHZ<br />

CALC:MARK:X:SLIM:RIGH 10MHZ<br />

CALC:MARK:X:SLIM ON<br />