R&S <strong>FSQ</strong> Instrument Functions<br />

Tracking Generator – Option R&S FSU-B9<br />

the power at the internal tracking generator is changed linear from start power to<br />

stop power. The start and stop power values are shown on the right side below the<br />

diagram.<br />

Remote command: :SOUR:POW:MODE SWE<br />

:SOUR:POW:MODE FIX<br />

START POWER The START POWER softkey defines the start power of the power sweep.<br />

The start power can be set between -30 dBm and +5 dBm.<br />

With the option R&S FSU-B12 the power can be set between -100 and + 5 dBm.<br />

Remote command: :SOUR:POW:STAR –20dBm<br />

STOP POWER The STOP POWER softkey defines the stop power of the power sweep.<br />

The start power can be set between -30 dBm and +5 dBm.<br />

With the option R&S FSU-B12 the power can be set between -100 and + 5 dBm.<br />

The stop value can be smaller than the start value.<br />

Remote command: :SOUR:POW:STOP –10dBm<br />

4.234 Operating Manual 1313.9681.12 - 01