R&S <strong>FSQ</strong> Remote Control – Description of Commands<br />

SYSTem:COMMunicate:GPIB:RDEVice:GENerator:ADDRess 0 to 30<br />

SYSTem Subsystem<br />

This command changes the GPIB address of the device selected as external generator 1 or 2.<br />

Example: "SYST:COMM:GPIB:RDEV:GEN1:ADDR 19"<br />

'Changes the GPIB address of generator 1 to 19<br />

Characteristics: *RST value: 28<br />

SCPI: device-specific<br />

Mode: A<br />

SYSTem:COMMunicate:GPIB[:SELF]:ADDRess 0 to 30<br />

This command changes the GPIB address of the unit.<br />

Example: "SYST:COMM:GPIB:ADDR 18"<br />

Characteristics: *RST value: - (no influence on this parameter, factory default 20)<br />

SCPI: conform<br />

Mode: A<br />

SYSTem:COMMunicate:GPIB[:SELF]:RTERminator LFEOI | EOI<br />

This command changes the GPIB receive terminator.<br />

According to the standard the terminator in ASCII is and/or . For binary data transfers (e.g.<br />

trace data) from the control computer to the instrument, the binary code (0AH) used for might be<br />

included in the binary data block, and therefore should not be interpreted as a terminator in this<br />

particular case. This can be avoided by changing the receive terminator to EOI.<br />

Output of binary data from the instrument to the control computer does not require such a terminator<br />

change.<br />

Example: "SYST:COMM:GPIB:RTER EOI"<br />

Characteristics: *RST value: -- (no influence on this parameter, factory default LFEOI)<br />

SCPI: device-specific<br />

Mode: A<br />

SYSTem:COMMunicate:PRINter:ENUMerate:FIRSt?<br />

This command queries the name of the first printer (in the list of printers) available under Windows NT.<br />

The names of other installed printers can be queried with command SYSTem:COMMunicate:<br />

PRINter:ENUMerate:NEXT?.<br />

If no printer is configured an empty string is output.<br />

Example: "SYST:COMM:PRIN:ENUM:FIRS?"<br />

Characteristics: *RST value: NONE<br />

SCPI: device-specific<br />

Mode: A<br />

If two generators are connected at the same time to IECBUS 2 their addresses<br />

must be different.<br />

The command is only available with option Ext. Generator Control, R&S FSP-B10.<br />

Operating Manual 1313.9681.12 - 01 6.227