R&S <strong>FSQ</strong> Remote Control – Description of Commands<br />

CALCulate Subsystem<br />

"CALC:MARK:FUNC:FPE 3"<br />

'searches the 3 highest maxima for trace 1<br />

"CALC:MARK:FUNC:FPE:COUN?"<br />

'queries the number of maxima found<br />

"CALC:MARK:FPE:FUNC:X?"<br />

'queries the frequencies (span 0) or time (span = 0) of the maxima found.<br />

Return value: "107.5E6,153.8E6,187.9E6"<br />

'frequencies in increasing order<br />

"2.05E-3,2.37E-3, 3.71e-3"<br />

'times in increasing order<br />

Characteristics: *RST value: --<br />

SCPI: device-specific<br />

Mode: A<br />

CALCulate:MARKer:FUNCtion:FPEaks:Y?<br />

This query reads out the list of X values of the maxima found. The number of available values can be<br />

queried with CALC:MARK:FUNC:FPEaks:COUNt?.<br />

With sort mode X, the X values are in increasing order; with sort mode Y the order corresponds to the<br />

decreasing order of the Y values.<br />

Example: "CALC:MARK:FUNC:FPE:SORT Y"<br />

'sets the sort mode to decreasing y values<br />

"CALC:MARK:FUNC:FPE 3"<br />

'searches the 3 highest maxima for trace 1<br />

"CALC:MARK:FUNC:FPE:COUN?"<br />

'queries the number of maxima found<br />

"CALC:MARK:FUNC:FPE:Y?"<br />

'queries the levels of the maxima found.<br />

Return value: "-37.5,-58.3,-59.6"<br />

'level in decreasing order<br />

Characteristics: *RST value: --<br />

SCPI: device-specific<br />

Mode: A<br />