18.12.2012 Views

R&S FSQ Signal Analyzer - Rohde & Schwarz

R&S FSQ Signal Analyzer - Rohde & Schwarz

R&S FSQ Signal Analyzer - Rohde & Schwarz

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

R&S <strong>FSQ</strong> Remote Control – Description of Commands<br />

HCOPy Subsystem<br />

The command HCOPy:DEVice:ITEM:WINDow:TABle:STATe OFF as well as command<br />

HCOPy:DEVice:ITEM:ALL enables the output of the whole screen.<br />

HCOPy:ITEM:WINDow:TEXT <br />

This command defines the comment text for measurement window 1 or 2 for printout, with a maximum<br />

of 100 characters; line feed by means of character @).<br />

Example: "HCOP:ITEM:WIND2:TEXT 'comment'"<br />

Characteristics: *RST value: -<br />

SCPI: device-specific<br />

Mode: A<br />

HCOPy:ITEM:WINDow:TRACe:STATe ON | OFF<br />

This command selects the output of the currently displayed trace.<br />

Example: "HCOP:ITEM:WIND:TRACe:STAT ON"<br />

Characteristics: *RST value: OFF<br />

SCPI: device-specific<br />

Mode: A<br />

The command HCOPy:ITEM:WINDow:TRACe:STATe OFF as well as command HCOPy:<br />

ITEM:ALL enables the output of the whole screen.<br />

HCOPy:PAGE:ORIentation LANDscape | PORTrait<br />

The command selects the format of the output (portrait and landscape) (hardcopy unit 1 or 2).<br />

Example: "HCOP:PAGE:ORI LAND"<br />

Characteristics: *RST value: -<br />

SCPI: conform<br />

Mode: A<br />

The command is only available provided that the output device "printer" (HCOP:<br />

DEST 'SYST:COMM:PRIN') has been selected.<br />

6.120 Operating Manual 1313.9681.12 - 01