R&S <strong>FSQ</strong> Instrument Functions<br />

4.8.2.2 Normalization<br />

External Generator Control – Option R&S FSP-B10<br />

NORMALIZE The NORMALIZE softkey switches normalization on or off. The softkey is only available<br />

if the memory contains a correction trace.<br />

It is possible to shift the relative reference point within the grid using the REF VALUE<br />

POSITION softkey. Thus, the trace can be shifted from the top grid margin to the<br />

middle of the grid:<br />

REF VALUE<br />

POSITION<br />

Fig. 4.30 Normalized display<br />

In the SPLIT SCREEN setting, the normalization is switched on in the current window.<br />

Different normalizations can be active in the two windows.<br />

Normalization is aborted when the NETWORK mode is quit.<br />

Remote command: CORR ON<br />

The REF VALUE POSITION softkey (reference position) marks a reference position<br />

in the active window on which the normalization (difference formation with a reference<br />

curve) is performed.<br />

When pressed for the first time, the softkey switches on the reference line and activates<br />

the input of its position. The line can be shifted within the grid limits.<br />

The reference line is switched off by pressing the softkey again.<br />

The function of the reference line is explained in the section “Calibration Mechanism”<br />

on page 4.243.<br />

Operating Manual 1313.9681.12 - 01 4.239