R&S <strong>FSQ</strong> Remote Control – Description of Commands<br />

6.16.3 SENSe:CORRection Subsystem<br />

SENSe Subsystem<br />

This subsystem also controls calibration and normalization during operation with the external generator<br />

control option (B10). The measurement windows are selected by SENSe1 (screen A) and SENSe2<br />

(screen B).<br />

[SENSe:]CORRection:COLLect[:ACQuire] THRough | OPEN<br />

When the tracking generator is active, this command determines the type of result acquisition for the<br />

normalization reference measurement and starts the measurement selected:<br />

Parameter: THRough:<br />

TRANsmission mode: calibration with direct connection between tracking<br />

generator and device input.<br />

REFLection mode: calibration with short circuit at the input<br />

OPEN: only valid in REFLection mode: calibration with open input<br />

Example: "INIT:CONT OFF"<br />

'Selects single sweep operation<br />

"CORR:COLL THR;*WAI"<br />

'Starts the measurement of reference data using direct connection between<br />

generator and device input and waits for the sweep end.<br />

Characteristics: *RST value: --<br />

SCPI: conform<br />

Mode: A<br />

This command is an "event" and therefore has no *RST value and no query.<br />

This command is only valid in conjunction with the ext. generator control option (B10).<br />

[SENSe:]CORRection:CVL:BAND A|Q|U|V|E|W|F|D|G|Y|J|USER<br />

This command defines the waveguide band in the Conversion Loss Table. Prior to this command, the<br />

command SENS:CORR:CVL:SEL must be sent.<br />

Example: "CORR:CVL:SEL 'LOSS_TAB'"<br />

"CORR:CVL:BAND E"<br />

Characteristics: *RST value: -<br />

SCPI: device-specific<br />

Mode: A<br />

This command is only available with the option R&S FSU-B21.<br />

[SENSe:]CORRection:CVL:BIAS <br />

This command defines the bias current in the Conversion Loss Table. Prior to this command, the<br />

command SENS:CORR:CVL:SEL must be sent.<br />

Example: "CORR:CVL:SEL 'LOSS_TAB'"<br />

"CORR:CVL:BIAS 7mA"<br />

Characteristics: *RST value: -<br />

SCPI: device-specific<br />

Mode: A<br />

This command is only available with the option R&S FSU-B21.<br />

Operating Manual 1313.9681.12 - 01 6.153