R&S <strong>FSQ</strong> Instrument Functions<br />

General Service Functions<br />

Basic Settings<br />

INPUT RF The INPUT RF softkey switches the input of the R&S <strong>FSQ</strong> to the input connector<br />

(normal position).<br />

After PRESET, RECALL or R&S <strong>FSQ</strong> power on, the INPUT RF is always selected.<br />

Remote command: DIAG:SERV:INP RF<br />

INPUT CAL The INPUT CAL softkey switches the RF input of the R&S <strong>FSQ</strong> to the internal calibration<br />

source (128 MHz) and activates the data entry of the output level of the calibration<br />

source. Possible values are 0 dB and –30 dB.<br />

ENTER<br />

PASSWORD<br />

CAL GEN 128<br />

MHZ<br />

CAL GEN<br />

COMB PULSE<br />

COMMAND<br />

TRACKING<br />

Remote command: DIAG:SERV:INP CAL;<br />

DIAG:SERV:INP:CSO 0 DBM<br />

The ENTER PASSWORD softkey allows the entry of a password.<br />

The R&S <strong>FSQ</strong> contains a variety of service functions which, if incorrectly used, can<br />

affect correct operation of the R&S <strong>FSQ</strong>. These functions are normally not accessible<br />

and are only usable after the entry of a password (see instrument service manual).<br />

Remote command: SYST:PASS "Password"<br />

The CAL GEN 128 MHZ softkey selects a sinusoidal signal at 128 MHz as output<br />

signal for the internal calibration source. The internal pulse generator will be<br />

switched off.<br />

CAL GEN 128 MHZ is the default setting of the R&S <strong>FSQ</strong>.<br />

Remote command: DIAG:SERV:INP CAL<br />

DIAG:SERV:INP:PULS OFF<br />

This softkey switches the internal pulse generator on and allows the pulse frequency<br />

to be entered.<br />

Available pulse frequencies are 10 kHz, 62,5 kHz, 1 MHz, 128 MHz, and 640 MHz.<br />

Remote command: DIAG:SERV:INP:PULS ON;<br />

DIAG:SERV:INP:PULS:PRAT <br />

The COMMAND TRACKING softkey activates or deactivates the SCPI error log<br />

function. All remote control commands received by the R&S <strong>FSQ</strong> are recorded in<br />

the following log file:<br />

D:\R_S\instr\log\IEC_CMDS.LOG<br />

Logging the commands may be extremely useful for debug purposes, e.g. in order<br />

to find misspelled keywords in control programs.<br />

If this softkey is activated, the R&S <strong>FSQ</strong> becomes slower.<br />

Remote command: --<br />

Operating Manual 1313.9681.12 - 01 4.201