R&S <strong>FSQ</strong> Remote Control – Description of Commands<br />

Characteristics: *RST value: --<br />

SCPI: device-specific<br />

Mode: A<br />

The numeric suffix at WINDow is irrelevant.<br />

DISPlay[:WINDow]:SELect<br />

DISPlay Subsystem<br />

This command selects the active measurement window. WINDow1 corresponds to SCREEN A,<br />

WINDow2 to SCREEN B.<br />

In full screen mode, the measurements are only performed in the active measurement window.<br />

Measurements are therefore initiated in the active window and result queries (marker, trace data and<br />

other results) answered also in the active window.<br />

Initiating measurements and querying results in the inactive window yields an error message<br />

(execution error).<br />

In split screen mode, the selection of the active window for result queries is irrelevant.<br />

In full screen mode, settings can also be performed in the inactive measurement window. They<br />

become effective as soon as the corresponding window becomes active.<br />

Example: "DISP:WIND2:SEL"<br />

'Selects screen B as active measurement window.<br />

Characteristics: *RST value: SCREEN A active<br />

SCPI: device-specific<br />

Mode: A<br />

This command is an event and therefore has no query.<br />

DISPlay[:WINDow]:SIZE LARGe | SMALl<br />

This command switches the measurement window for channel and adjacent channel power<br />

measurements to full screen or half screen. Only "1" is allowed as a numerical suffix.<br />

Example: "DISP:WIND1:SIZE LARG"<br />

'Switches the measurement window to full screen.<br />

Characteristics: *RST value: SMALl<br />

SCPI: device-specific<br />

Mode: A<br />

DISPlay[:WINDow]:TEXT[:DATA] <br />

This command defines a comment (max. 20 characters) which can be displayed on the screen in the<br />

selected measurement window.<br />

Example: "DISP:WIND2:TEXT 'Noise Measurement'"<br />

'Defines the title for screen B<br />

Characteristics: *RST value: "" (empty)<br />

SCPI: conform<br />

Mode: A<br />

DISPlay[:WINDow]:TEXT:STATe ON | OFF<br />

This command switches on or off the display of the comment (screen title) in the selected<br />

measurement window.<br />