R&S <strong>FSQ</strong> Instrument Functions<br />

COUPLING<br />

RATIO<br />

RBW / VBW<br />

SINE [1/3]<br />

RBW / VBW<br />

PULSE [0.1]<br />

<strong>Analyzer</strong> Mode<br />

The SWEEPTIME AUTO function is designed for detectors like sample, max peak or<br />

average. For sweeping with the quasipeak detector a manual setting of the sweep<br />

time should be used at any time.<br />

Remote command: SWE:TIME:AUTO ON<br />

The COUPLING RATIO softkey opens a submenu for selection of the coupling ratios.<br />

RBW / VBW SINE [1/3]<br />

RBW / VBW PULSE [0.1]<br />

RBW / VBW NOISE [10]<br />

RBW / VBW MANUAL<br />

SPAN / RBW AUTO [50]<br />

SPAN / RBW MANUAL<br />

When the default setting is active, i.e. the COUPLING RATIO softkey is deactivated<br />

(not highlighted), the ratio span/resolution bandwidth (SPAN/RBW) is 50 (this corresponds<br />

to SPAN / RBW AUTO [50]) and the ratio resolution bandwidth/video bandwidth<br />

(RBW/VBW) is 0.33 (this corresponds to RBW / VBW SINE [1/3]).<br />

If the ratio RBW/VBW or SPAN/RBW is different from the default setting, the COU-<br />

PLING RATIO softkey is highlighted.<br />

The softkeys RBW/VBW... are selection keys. Only one softkey can be enabled at<br />

any one time. The softkeys are only effective for the VBW AUTO selection in the<br />

main menu.<br />

The softkeys SPAN/RBW... are also selection keys. They are only effective for the<br />

RBW AUTO selection in the main menu.<br />

The RBW / VBW SINE [1/3] softkey sets the following coupling ratio:<br />

video bandwidth = 3 × resolution bandwidth.<br />

This is the default setting for the coupling ratio resolution bandwidth/video bandwidth.<br />

This is the coupling ratio recommended if sinusoidal signals are to be measured.<br />

Remote command: BAND:VID:RAT 3<br />

This setting is only effective for the VBW AUTO selection in the main menu.<br />

The RBW / VBW PULSE [0.1] softkey sets the following coupling ratio:<br />

video bandwidth = 10 × resolution bandwidth or<br />

video bandwidth = 10 MHz (= max. VBW).<br />

This coupling ratio is recommended whenever the amplitudes of pulsed signals are<br />

to be measured correctly. The IF filter is exclusively responsible for pulse shaping.<br />

No additional evaluation is performed by the video filter.<br />

Remote command: BAND:VID:RAT 10<br />

This setting is only effective for the VBW AUTO selection in the main menu.<br />