R&S <strong>FSQ</strong> Instrument Functions<br />

THRESHOLD The THRESHOLD softkey defines the threshold line.<br />

SEARCH LIMIT<br />

OFF<br />

<strong>Analyzer</strong> Mode<br />

The threshold line represents a limit for the level range of the max. search at the<br />

lower end and that of the min. search at the upper end.<br />

Remote command: CALC:THR -20dBm<br />

CALC:THR ON<br />

The SEARCH LIMIT OFF softkey disables all limits of the search range.<br />

Remote command: CALC:MARK:X:SLIM OFF<br />

CALC:THR OFF<br />

MKR->TRACE The function of this softkey is identical to that of the MKR->TRACE softkey in the<br />

MKR menu (see “MKR->TRACE” on page 4.59).<br />

MKR->CF<br />

STEPSIZE<br />

The MKR->CF STEPSIZE softkey sets the step size for the center frequency variation<br />

to the current marker frequency, and also sets step size adaptation to MANUAL.<br />

CF STEPSIZE remains at this value until the center frequency entry mode in the<br />

STEP menu is switched from MANUAL to AUTO again.<br />

The MKR->CF STEPSIZE function is, above all, useful in the measurement of harmonics<br />

with large dynamic range (narrow bandwidth and narrow span).<br />

The softkey is not available in the time domain (span = 0 Hz).<br />

Remote command: CALC:MARK:FUNC:CST<br />

Example<br />

The harmonics levels of a CW carrier are to be measured at 100 MHz.<br />

[PRESET] R&S <strong>FSQ</strong> is set to the default setting.<br />

[CENTER: 100 MHz] R&S <strong>FSQ</strong> sets the center frequency to 100 MHz. The span<br />

is set to 200 MHz.<br />

[SPAN: 1 MHz] The span is set to 100 MHz.<br />

[MKR->] MARKER 1 is switched on and set to the maximum value<br />

of the signal.<br />

[NEXT] R&S <strong>FSQ</strong> switches to the submenu.<br />

[MKR->CF STEPSIZE] The step size of the center frequency setting equals the<br />

marker frequency (100 MHz).<br />

[CENTER] The center frequency entry mode is activated.<br />

[Right key] The center frequency is set to 200 MHz. The first harmonic<br />

of the test signal is displayed.<br />

[MKR->: PEAK] The marker is set to the harmonic and the level of the latter<br />

is output in the marker info field.<br />

MIN The MIN softkey sets the active marker/delta marker to the minimum of the selected<br />

trace.<br />

Remote command: CALC:MARK:MIN<br />

CALC:DELT:MIN<br />

Operating Manual 1313.9681.12 - 01 4.75