R&S <strong>FSQ</strong> Instrument Functions<br />

CAL CORR<br />

(ON OFF)<br />

YIG CORR<br />

(ON OFF)<br />

The CAL CORR (ON OFF) softkey switches the calibration data on/off.<br />

<strong>Analyzer</strong> Mode<br />

• ON: The status message depends upon the results of the total calibration.<br />

• OFF: The message UNCAL appears in the status line.<br />

Remote command: CAL:STAT ON<br />

The YIG CORR (ON OFF) softkey switches on or off the automatic, cyclic correction<br />

of the temperature-dependent frequency drift of the YIG filter.<br />

The YIG CORR (ON OFF) softkey is available only on models with a frequency<br />

range ≥ 3 GHz and ≤ 40 GHz.<br />

When switched to ON (default setting), it is checked once per minute whether a frequency<br />

correction for the YIG filter is required. Frequency correction is performed if<br />

the temperature has changed by more than 3K relative to the last instance of correction.<br />

If the instrument is operated in a temperature-controlled environment, the cyclic<br />

frequency drift correction can – for time-critical applications – be switched off after<br />

an operating period of ≥ 30 minutes.<br />

Remote command: CORR:YIG:TEMP:AUTO ON | OFF<br />

CAL RESULTS The CAL RESULTS softkey calls the CALIBRATION RESULTS table, which shows<br />

the correction data found during calibration.<br />

The CALIBRATION RESULTS table contains the following information:<br />

– date and time of last record of correction values<br />

– overall results of correction value record<br />

– list of found correction values according to function/module<br />

The results have the following meaning:<br />

PASSED calibration successful without any restrictions<br />

CHECK deviation of correction value larger than expected, correction<br />

could however be performed<br />

FAILED deviations of correction value too large, no correction was possible.<br />

The found correction data are not valid.<br />

ABORTED calibration aborted<br />

Operating Manual 1313.9681.12 - 01 4.53