R&S <strong>FSQ</strong> Instrument Functions<br />

CAL REFL<br />

OPEN<br />

The message is cleared after approx. 3 seconds.<br />

Remote command: CORR:METH REFL<br />

CORR:COLL THR<br />

Tracking Generator – Option R&S FSU-B9<br />

The CAL REFL OPEN softkey starts the open-circuit calibration. During calibration<br />

the following message is displayed:<br />

Remote command: CORR:METH REFL<br />

CORR:COLL OPEN<br />

4.7.4 Calibration Mechanism<br />

Calibration means a calculation of the difference between the currently measured<br />

power and a reference curve, independent of the selected type of measurement<br />

(transmission/reflection). The hardware settings used for measuring the reference<br />

curve are included in the reference data set.<br />

Even with normalization switched on, the device settings can be changed in a wide<br />

area without stopping the normalization. This reduces the necessity to carry out a<br />

new normalization to a minimum.<br />

For this purpose the reference dataset (trace with 625 measured values) is stored<br />

internally as a table of 625 points (frequency/level).<br />

Differences in level settings between the reference curve and the current device settings<br />

are taken into account automatically. If the span is reduced, a linear interpolation<br />

of the intermediate values is applied. If the span increases, the values at the left<br />

or right border of the reference data set are extrapolated to the current start or stop<br />

frequency, i.e. the reference data set is extended by constant values.<br />

An enhancement label is used to mark the different levels of measurement accuracy.<br />

This enhancement label is displayed at the right diagram border when normalization<br />

is switched on and a deviation from the reference setting occurs. Three<br />

accuracy levels are defined:<br />

Table 4-6 Measurement accuracy levels<br />

Accuracy Enhancement<br />

label<br />

Reason/Limitation<br />

High NOR No difference between reference setting and<br />

measurement<br />

Operating Manual 1313.9681.12 - 01 4.229