R&S <strong>FSQ</strong> Remote Control – Description of Commands<br />

OUTPut[:STATe] ON | OFF<br />

This command switches the tracking generator on or off.<br />

Example: "OUTP ON"<br />

'Switches on the tracking generator in screen A.<br />

Characteristics: *RST value: OFF<br />

SCPI: conform<br />

Mode: A<br />

OUTPut Subsystem<br />

• With the tracking generator switched on, the maximum stop frequency is limited<br />

to 3 GHz. This upper limit is automatically modified by the set frequency offset<br />

of the generator.<br />

• If measurements in compliance with specs are to be performed with the tracking<br />

generator, the start frequency has to be ≥3 x resolution bandwidth.<br />

• The minimum sweep time for measurements in compliance with the data sheet<br />

is 100 ms in the frequency domain (span >0). If a shorter sweep time is<br />

selected, the sweep time indicator SWT on the screen is marked with a red<br />

asterisk and the message UNCAL is also displayed.<br />

• With the tracking generator switched on, the FFT filters (BAND:MODE:FFT) are<br />

not available.<br />

This command is only valid in conjunction with option tracking generator R&S FSU-B9.<br />

Operating Manual 1313.9681.12 - 01 6.145