R&S <strong>FSQ</strong> Remote Control – Description of Commands<br />

6.5.4 CALCulate:FLINe Subsystem<br />

CALCulate Subsystem<br />

The CALCulate:FLINe subsystem controls the settings for the frequency limit lines. The measurement<br />

windows are selected by CALCulate 1 (screen A) or 2 (screen B).<br />

CALCulate:FLINe 0 to fmax<br />

This command defines the position of the frequency lines.<br />

The frequency lines mark the frequencies specified in the measurement window. Frequency lines are<br />

only available with SPAN > 0.<br />

Example: "CALC:FLIN2 120MHz"<br />

Characteristics: *RST value: - (STATe to OFF)<br />

SCPI: device-specific<br />

Mode: A-F<br />

CALCulate:FLINe:STATe ON | OFF<br />

This command switches the frequency line on or off.<br />

Example: "CALC:FLIN2:STAT ON"<br />

Characteristics: *RST value: OFF<br />

SCPI: device-specific<br />

Mode: A-F<br />