18.12.2012 Views

R&S FSQ Signal Analyzer - Rohde & Schwarz

R&S FSQ Signal Analyzer - Rohde & Schwarz

R&S FSQ Signal Analyzer - Rohde & Schwarz

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

R&S <strong>FSQ</strong> Remote Control – Description of Commands<br />

6.6 CALibration Subsystem<br />

CALibration Subsystem<br />

The commands of the CALibration subsystem determine the data for system error correction in the<br />

instrument.<br />

CALibration:ABORt<br />

This command aborts the acquisition of correction data and restores the last complete correction data<br />

set.<br />

Example: "CAL:ABOR"<br />

Characteristics: *RST value: -<br />

SCPI: device-specific<br />

Mode: A<br />

This command is an event and therefore has no *RST value and no query.<br />

CALibration[:ALL]?<br />

This command initiates the acquisition of system error correction data. A "0" is returned if the<br />

acquisition was successful.<br />

In order to recognize when the acquisition of correction data is completed, the MAV bit in the status<br />

byte can be used. If the associated bit is set in the Service Request Enable Register, the instrument<br />

generates a service request after the acquisition of correction data has been completed.<br />

Example: "*CLS"<br />

'Resets the status management.<br />

Characteristics: *RST value: -<br />

SCPI: conform<br />

Mode: A<br />

CALibration:RESult?<br />

During the acquisition of correction data, the instrument does not accept any<br />

remote control commands, except<br />

*RST<br />

CALibration:ABORt<br />

"*SRE 16"<br />

'Enables MAV bit in the Service Request Enable Register.<br />

"*CAL?"<br />

'Starts the correction data recording and then a service request is generated.<br />

This command outputs the results of the correction data acquisition. The lines of the result table (see<br />

chapter “Instrument Functions”, section “Recording the Correction Data – CAL Key” on page 4.52) are<br />

output as string data separated by commas:<br />

Return value: "Total Calibration Status: PASSED","Date (dd/mm/yyyy): 12/07/<br />

2006",<br />

"Time: 16:24:54","Runtime:00.06"<br />

Example: "CAL:RES?"<br />

Operating Manual 1313.9681.12 - 01 6.101

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!