R&S <strong>FSQ</strong> Remote Control – Description of Commands<br />

STATus:OPERation:PTRansition 0 to 65535<br />

STATus Subsystem<br />

This command sets the edge detectors of all bits of the STATus:OPERation register from 0 to 1 for the<br />

transitions of the CONDition bit.<br />

Example: "STAT:OPER:PTR 65535"<br />

Characteristics: *RST value: –<br />

SCPI: conform<br />

Mode: A<br />

STATus:PRESet<br />

This command resets the edge detectors and ENABle parts of all registers to a defined value. All<br />

PTRansition parts are set to FFFFh, i.e. all transitions from 0 to 1 are detected. All NTRansition parts<br />

are set to 0, i.e. a transition from 1 to 0 in a CONDition bit is not detected. The ENABle part of the<br />

STATus:OPERation and STATus:QUEStionable registers are set to 0, i.e. all events in these registers<br />

are not passed on.<br />

Example: "STAT:PRES"<br />

Characteristics: *RST value: –<br />

SCPI: conform<br />

Mode: A<br />

STATus:QUEStionable:ACPLimit:CONDition?<br />

This command queries the contents of the CONDition section of the STATus:QUEStionable:ACPLimit<br />

register. Readout does not delete the contents of the CONDition section.<br />

Return value: see “STATus:QUEStionable:ACPLimit Register” on page 5.27<br />

Example: "STAT:QUES:ACPL:COND?"<br />

Characteristics: *RST value: –<br />

SCPI: device-specific<br />

Mode: A<br />

STATus:QUEStionable:ACPLimit:ENABle 0 to 65535<br />

This command sets the bits of the ENABle section of the STATus:QUEStionable:ACPLimit register.<br />

The ENABle register selectively enables the individual events of the associated EVENt section for the<br />

summary bit.<br />

Example: "STAT:QUES:ACPL:ENAB 65535"<br />

Characteristics: *RST value: –<br />

SCPI: device-specific<br />

Mode: A<br />

Operating Manual 1313.9681.12 - 01 6.219